GE PT878GC Manual de usuario Pagina 64

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 308
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 63
Programming Site Data 4-3
April 2004
Entering Transducer Parameters
To enter the Transducer option, scroll to the Transducer entry on
the Program Menu and press
[ENTER]. The screen appears
similar to Figure 4-2 below. To step through each parameter (or
reach a particular parameter), press the [
T] key. Refer to
Figure A-3 on page A-3 of Appendix A, Menu Maps
Note: Refer to Chapter 3 for additional information about
clamp-on gas transducers and fixture configurations.
.
Figure 4-2: The Transducer Option Window
1. The first prompt asks you to select whether you are using a
wetted or a clamp-on transducer.
a. Use the [
W] and [X] keys to scroll between the two types.
b. Press
[ENTER] or [SEL] to confirm the choice.
IMPORTANT: While wetted transducers are listed as a choice,
the PT878GC is designed for use with Shear
wave clamp-on transducers. Consult GE
Panametrics before selecting wetted transducers.
Transducer
Pipe
Lining Path
Transducer/Pipe
OK
Cancel
Fluid
Type: Wetted
Clamp-on
Transducer
Special
Frequency
2.00
Tw
14
MHz
Wedge Ang
50
°
Wedge Tmp
25
°C
Wedge SS
1219.2
m/s
µs
Vista de pagina 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 307 308

Comentarios a estos manuales

Sin comentarios