9-22 Servicing the PT878GC
Calibrating the Analog Output (cont.)
1. Connect a digital multimeter or ammeter to the analog output.
2. The next prompt asks for the setpoint (4 or 20 mA). Use the
[
W ] and [X] arrow keys to move to the appropriate radio
button and press
[ENTER].
3. The next prompt asks you to enter the actual value shown on
the multimeter or ammeter.
a. Press
[ENTER] to open the text box.
b. Use the numeric keys to enter the measured value.
c. Press
[ENTER] to confirm your entry.
4. Repeat steps 2 and 3 for the other setpoint.
5. Press the [
T ] arrow key to move to the Calibrate box, and
then press
[ENTER] to calibrate the output.
If you are unsatisfied with the calibration, you can reset the inputs
to factory defaults. Press the [
X] arrow key to move to the Reset
box, and then press
[ENTER] to undo the calibration.
You have completed calibrating the outputs. To leave the Output
window,
Press the [S] key to return to the Output tab, and the [X]
arrow key to move to the Input tab.
Press [F2] (Cancel) to return to Operate Mode without
confirming the changes.
Press [F3] (OK) to confirm the new data. The PT878GC
returns to Operate Mode.