GE PT878GC Manual de usuario Pagina 240

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 308
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 239
Servicing the PT878GC 9-39
April 2004
Setting Impulse Response
The Impulse Response option enables you to force the meter to
transmit in one direction only, without changing the AGC setting.
You can then diagnose problems with transducer, pipe, or fluid
configurations. To enter the option, scroll down to the Impulse
Response entry in the Service Menu and press
[ENTER]. The
screen appears similar to Figure 9-24 below.
Figure 9-24: The Impulse Response Option
1. The first prompt asks in which direction you wish the meter to
transmit, upstream or downstream.
a. Press
[ENTER] to open the drop-down list.
b. Use the [
T] or [S] arrow keys to scroll to the desired
direction.
c. Press
[ENTER] to confirm your selection.
Transmit
Stop Exit
Impulse Response
.
Direction Upstream
Percent Gain
20 %
Stopped.
Vista de pagina 239
1 2 ... 235 236 237 238 239 240 241 242 243 244 245 ... 307 308

Comentarios a estos manuales

Sin comentarios